Analytical Specifications | DNA 1K Kit |
---|---|
Sizing Range | 0.1-1.5 kb |
Limit of Detection (LOD) (1) | 10 pg/mL at 1kb 50 pg/mL at 200 bp |
Sizing Accuracy | ± 3% |
Sizing Reproducibility | < 3% CV |
Quantitative Range for DNA Fragments | 40 pg/mL to 10 ng/mL |
Quantitative Range for Biological Samples | 5 – 1000 ng/mL for cfDNA |
Quantification Accuracy | ± 20% |
Quantification Precision | < 20% CV |
Sample Volume | 10 µL (1 µL injected) |
Salt Concentration (2) | 0-15 mM for LoSalt method 0-130 mM for HiSalt method |
Turn Around Time (LoSalt method) | 37 minutes |
Kit Stability | 2 years |
1 Limit of detection : concentration for that signal to noise ratio (SNR) equals 3. Signal : height of the fragment peak. Noise : peak-to-peak noise measured on the baseline. | ||
2 For NaCl in TE |